Abstract:
Objective To study the dose distribution of
125Ⅰ seeds in experiment phantom.
Methods The dose distribution of
125Ⅰ seeds was performed by the thermolnminescence dosimetry (TLD) method and the dose calculation formula (calculation). The radial dose of
125Ⅰ seed was compared using measured and calculated result.
Results When radioactivity of
125Ⅰ seed was 37MBq (1.OmCi) and the radial distance is 0.5, 1.0. 1.5 and 2cm, the TLD measured and theoretically calculated result is 3.47, 0.87, 0.38, 0.21 eGy/h and 3.874, 0.952, 0.394, 0.216 eGy/h respectively.
Conclusions The TLD measured and theoretically calculated result were found to good agreement. The radial dose of
125Ⅰ seed is attenuated very apidly with radial distance.