影像设备与放疗技术发展的关系越来越紧密,而包含电子射野影像系统(electronic portal imaging device, EPID)的加速器已在放疗中得到了广泛使用,EPID始于20世纪80年代,是使用荧光显像系统获得兆伏级传输图像的技术,在放疗质量控制和质量保证中应用广泛:图像方面如摆位误差验证分析等,加速器自身方面如射野、等中心验证、多叶准直器位置和到位精度验证等,剂量方面如二维三维剂量验证等,同时也有部分用于实时图像引导和剂量引导放疗的尝试。近年来,EPID设备和相关软件发展更新迅速,本文就EPID在放疗中的应用研究进展进行综述。
电子射野影像系统在放疗中的应用进展
Advances in the application of electronic portal imaging device in radiotherapy
-
摘要: 电子射野影像系统已成为放疗质量控制和质量保证的重要设备之一,可用于摆位误差验证分析、加速器本身日常质量控制、剂量验证、实时剂量验证等放疗质控指标的采集分析,以保障放疗实施的准确性。笔者就电子射野影像系统在放疗中的应用研究进行简要综述。Abstract: Electronic portal imaging devices have become important equipment for radiotherapy quality control and quality assurance. They can be used to verify positioning errors and other quality control indicators. They can also be applied in the daily quality control of the accelerator, dosimetry, and real-time dosimetry to ensure the accuracy of radiotherapy implementation. In this paper, the application of electronic portal imaging devices in radiotherapy is briefly reviewed.
-
Key words:
- Radiotherapy /
- Quality control /
- Electronic portal imaging device /
- Online verification
-
[1] Estoesta RP, Attwood L, Naehrig D, et al. Assessment of voluntary deep inspiration breath-hold with CINE imaging for breast radiotherapy[J]. J Med Imaging Radiat Oncol, 2017, 61 (5):689-694. DOI:10.1111/1754-9485.12616. [2] Li XA, Chen X, Zhang Q, et al. Margin Reduction from IGRT for Soft-Tissue Sarcoma: Secondary Analysis of RTOG0630 Results[J/OL]. Pract Radiat Oncol, 2016, 6 (4): e135-e140[2018-01-10]. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4870138. DOI: 10.1016/j.prro.2015.11.012. [3] Held M, Cremers F, Sneed PK, et al. Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments[J]. J Appl Clin Med Phys, 2016, 17 (2):279-290. DOI:10.1120/jacmp.v17i2.6040. [4] Held M, Sneed PK, Fogh SE, et al. Feasibility of MV CBCT-based treatment planning for urgent radiation therapy:dosimetric accuracy of MV CBCT-based dose calculations[J]. J Appl Clin Med Phys, 2015, 16 (6):458-471. DOI:10.1120/jacmp.v16i6.5625. [5] Poels K, Verellen D, Van de Vondel I, et al. Fiducial marker and marker-less soft-tissue detection using fast MV fluoroscopy on a new generation EPID: investigating the influence of pulsing artifacts and artifact suppression techniques[J/OL]. Med Phys, 2014, 41 (10): 101911[2018-01-10]. https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.4896116. DOI: 10.1118/1.4896116. [6] Bryant JH, Rottmann J, Lewis JH, et al. Registration of clinical volumes to beams-eye-view images for real-time tracking[J/OL]. Med Phys, 2014, 41 (12): 121703[2018-01-10]. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4235653. DOI: 10.1118/1.4900603. [7] Rottmann J, Aristophanous M, Chen A, et al. A multi-region algorithm for markerless beam's-eye view lung tumor tracking[J]. Phy Med Biol, 2010, 55 (18):5585-5598. DOI:10.1088/0031-9155/55/18/021. [8] Cho B, Poulsen PR, Sloutsky A, et al. First demonstration of combined kV/MV image-guided real-time dynamic multileaf-collimator target tracking[J]. Int J Radiat Oncol Biol Phys, 2009, 74 (3):859-867. DOI:10.1016/j.ijrobp.2009.02.012. [9] Rottmann J, Keall P, Berbeco R. Markerless EPID image guided dynamic multi-leaf collimator tracking for lung tumors[J]. Phys Med Biol, 2013, 58 (12):4195-5204. DOI:10.1088/0031-9155/58/12/4195. [10] Chiu TD, Yan Y, Foster R, et al. Long-term evaluation and cross-checking of two geometric calibrations of kV and MV imaging systems for Linacs[J]. J Appl Clin Med Phys, 2015, 16 (4):306-310. DOI:10.1120/jacmp.v16i4.5140. [11] Buzurovic I, Huang K, Yu Y, et al. A robotic approach to 4D real-time tumor tracking for radiotherapy[J]. Phys Med Biol, 2011, 56 (5):1299-1318. DOI:10.1088/0031-9155/56/5/005. [12] Sun B, Goddu SM, Yaddanapudi S, et al. Daily QA of linear accelerators using only EPID and OBI[J]. Med Phys, 2015, 42 (10):5584-5594. DOI:10.1118/1.4929550. [13] Rowshanfarzad P, Riis HL, Zimmermann SJ, et al. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation[J/OL]. Br J Radiol, 2015, 88 (1051): 20140581[2018-01-10]. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4628515. DOI: 10.1259/bjr.20140581. [14] Huang YC, Yeh CY, Yeh JH, et al. Clinical practice and evaluation of electronic portal imaging device for VMAT quality assurance[J]. Med Dosim, 2013, 38 (1):35-41. DOI:10.1016/j.meddos.2012. 05. 004. [15] 黎旦, 宾石珍, 程品晶, 等.非晶硅电子射野影像装置在宫颈癌剂量验证中的应用[J].中国医学物理学杂志, 2017, 34 (3):230-234. DOI:10.3969/j.issn.1005-202X.2017.03.003.
Li D, Bin SZ, Cheng PJ, et al. Dose verification of cervical cancer using a-Si EPID[J]. Chin J Med Phys, 2017, 34 (3):230-234. doi: 10.3969/j.issn.1005-202X.2017.03.003[16] Woodruff HC, Fuangrod T, Rowshanfarzad P, et al. Gantry-angle resolved VMAT pretreatment verification using EPID image prediction[J/OL]. Med Phys, 2013, 40 (8): 081715[2018-01-10]. https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.4816384.DOI:10.1118/1.4816384" target="_blank">10.1118/1.4816384">https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.4816384.DOI:10.1118/1.4816384. [17] Min S, Choi YE, Kwak J, et al. Practical approach for pretreatment verification of IMRT with flattening filter-free (FFF) beams using Varian Portal Dosimetry[J/OL]. J Appl Clin Med Phys, 2014, 16 (1): 4934[2018-01-10]. https://www.ncbi.nlm.nih.gov/pubmed/?term=25679149. DOI: 10.1120/jacmp.v16i1.4934. [18] Faught AM, Wu Q, Adamson J. SU-F-T-283:A Novel Device to Enable Portal Dosimetry for Flattening Filter Free Beams[J]. Med Phys, 2016, 43 (6):3527-3528. DOI:10.1118/1.4956423. [19] Pardo E, Novais JC, Molina López MY, et al. On flattening filter-free portal dosimetry[J]. J Aappl Clin Medl Phys, 2016, 17 (4):132-145. DOI:10.1120/jacmp.v17i4.6147. [20] Chuter RW, Rixham PA, Weston SJ, et al. Feasibility of portal dosimetry for flattening filter-free radiotherapy[J]. J Applied Clin Med Phys, 2016, 17 (1):112-120. DOI:10.1120/jacmp.v17i1.5686. [21] Willett A, Gilmore M, Rowbottom C. SU-F-T-567:Sensitivity and Reproducibility of the Portal Imaging Panel for Routine FFF QC and Patient Plan Dose Measurements[J]. Med Phys, 2016, 43 (6):3593-3594. DOI:10.1118/1.4956752. [22] Miri N, Keller P, Zwan BJ, et al. EPID-based dosimetry to verify IMRT planar dose distribution for the aS1200 EPID and FFF beams[J]. J Appl Clin Med Phys, 2016, 17 (6):292-304. DOI:10.1120/jacmp.v17i6.6336. [23] Mijnheer BJ, González P, Olaciregui-Ruiz I, et al. Overview of 3-year experience with large-scale electronic portal imaging device-based 3-dimensional transit dosimetry[J/OL]. Pract Radiat Oncol, 2015, 5 (6): e679-e687[2018-01-10]. https://www.sciencedirect.com/science/article/pii/S1879850015002313?via%3Dihub. DOI: 10.1016/j.prro.2015.07.001. [24] Fidanzio A, Porcelli A, Azario L, et al. Quasi real time in vivo dosimetry for VMAT[J/OL]. Med Phys, 2014, 41 (6): 062103[2018-01-10]. https://aapm.onlinelibrary.wiley.com/doi/abs/10.1118/1.4875685. DOI: 10.1118/1.4875685. [25] Spreeuw H, Rozendaal R, Olaciregui-Ruiz I, et al. Online 3D EPID-based dose verification:Proof of concept[J]. Med Phys, 2016, 43 (7):3969-3974. DOI:10.1118/1.4952729. [26] McCowan PM, Van Uytven E, Van Beek T, et al. An in vivo dose verification method for SBRT-VMAT delivery using the EPID[J]. Med Phys, 2015, 42 (12):6955-6963. DOI:10.1118/1.4935201. [27] Millin AE, Windle RS, Lewis DG. A comparison of electronic portal dosimetry verification methods for use in stereotactic radiotherapy[J]. Phys Med, 2016, 32 (1):188-196. DOI:10.1016/j.ejmp.2015. 12.001. [28] Fuangrod T, Greer PB, Woodruff HC, et al. Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits[J/OL]. Radiat Oncol, 2016, 11 (1): 106[2018-01-10]. https://ro-journal.biomedcentral.com/articles/10.1186/s13014-016-0682-y. DOI: 10.1186/s13014-016-0682-y. [29] Woodruff HC, Fuangrod T, Van Uytven E, et al. First Experience With Real-Time EPID-Based Delivery Verification During IMRT and VMAT Sessions[J]. Int J Radiat Oncol Biol Phys, 2015, 93 (3):516-522. DOI:10.1016/j.ijrobp.2015.07.2271. [30] Yip S, Rottmann J, Berbeco R. The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking[J/OL]. Med Phys, 2014, 41 (6): 061702[2018-01-10]. https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4032434. DOI: 10.1118/1.4873322. [31] Wang X, Chen L, Xie C, et al. Experimental verification of a 3D in vivo dose monitoring system based on EPID[J]. Oncotarget, 2017, 8 (65):109619-109631. DOI:10.18632/oncotarget.22758.
计量
- 文章访问数: 3658
- HTML全文浏览量: 2571
- PDF下载量: 7