调强适形放射治疗(intensity modulated radiation therapy,IMRT)是20世纪90年代发展起来的一项先进放射治疗技术,它具有靶区的高剂量区三维适形、周围危及器官受量少等优点,能有效提高肿瘤的局部控制率,减少正常组织的损伤,改善患者的生存质量。由于IMRT是一项复杂的技术,为保证其临床治疗的安全可靠,必须对IMRT计划进行准确的剂量验证。
放射治疗计划系统依据CT电子密度体模确认的CT密度转换曲线,将CT图像中的CT值转换为相应的物理密度(或电子密度),在物理密度(或电子密度)的基础上进行组织不均匀性校正计算,最终得到放射治疗计划的剂量分布[1-2]。然而,由于刻度CT密度转换曲线的CT电子密度体模缺乏与IMRT验证体模密度相接近的组织材料,因此,需要计划系统依据转换曲线的邻近区间,插值计算出IMRT验证体模的物理密度。这插值计算出来的物理密度与实际验证体模的密度存在少许差异,而这些差异将在一定程度上影响IMRT计划验证的精确性。本研究通过实验,定量分析CT密度转换曲线在IMRT剂量验证中的影响,并加以校准。
-
(1)美国CIRS公司Model 062电子密度体模,为27 cm×33 cm×5 cm的椭圆型薄柱体,环氧树脂材料,含有人体等效组织材料9种,几何参数及各插件位置、尺寸、密度参数如图 1和表 1所示。
物理密度(g/cm3) 电子密度(x1023/m3) 相对水电子密度 体模(水等效) 1.01 3.346 1.002 肺(吸气) 0.20 0.634 0.190 肺(呼气) 0.50 1.632 0.489 脂肪 0.96 3.170 0.949 乳腺 0.99 3.261 0.976 肌肉 1.06 3.483 1.043 肝 1.07 3.516 1.052 骨(200mg/cm3) 1.16 3.730 1.117 致密骨(800 mg/cm3) 1.53 4.862 1.456 表 1 美国CIRS公司Model 062电子密度体模的密度参数表
(2)德国IBA公司IMRT验证体模,由各种不同尺寸的模块构成,各模块具有规矩的形状,为正方体或长方体。整个体模均由同一种材料RW3(一种等效水材料)组成,主要成分为C8H8,含2.1% ± 0.2%的TiO2。
(3)德国IBA公司的二维电离室矩阵MatriXX,由1020个通气的电离室排列成32×32平面矩阵(四个顶角处无电离室)。电离室尺寸为直径4.5 mm、高5 mm,灵敏体积0.07 cm3,相邻电离室中心间距为7.62 mm,电离室矩阵有效测量面积为24.4 cm×24.4 cm,有效测量点位于上表面下3 mm。每个电离室在出厂时已经用60Co γ射线进行校准,刻度因子以矩阵文件形式给出。
(4)飞利浦公司的放射治疗计划系统Pinacle,版本为8.0。美国Picker公司的螺旋CT机,型号为PQ5000,定期由工程师负责校准检测。美国Varian公司的医用直线加速器23EX,测量前均检测、校准机器机械参数、输出剂量等,确保机器性能稳定。德国IBA公司的指形电离室FC65-G,灵敏体积0.65 cm3;德国IBA公司的剂量仪DOSE-1, 二者均由广东省辐射剂量计量检定站进行检定。
-
在螺旋CT机下扫描CT电子密度体模,扫描床板为放疗专用碳素平板,扫描电压为120 kV,电流为200 mA,扫描方式为螺旋扫描: 层厚3 mm,层距3 mm。将图像传至计划系统,勾画出体模中各插件,由计划系统计算出材料的平均CT值。将各等效组织材料的平均CT值与对应的已知物理密度值输入计划系统Pinacle,绘制出相应的CT密度转换曲线。
-
随机选取12例鼻咽癌患者的静态IMRT计划,计划参数为7个照射野,全周等分,最大子野数为90个,子野数范围为66~90个,计划设计中滤去面积小于4 cm2以及机器跳数小于5 MU(计划系统、加速器上使用的剂量计算单位,1 MU=1 cGy)的小子野。
-
在同一CT机下,以相同的扫描条件扫描IMRT验证体模,图像传输至治疗计划系统,三维重建该验证体模。将12例鼻咽癌IMRT计划分别移植至IMRT验证体模中,按照计划的实际照射角度,计算剂量,剂量以机器跳数归一,于靶区剂量均匀处(剂量梯度变化小于±6%)放置测量点。
剂量验证在加速器23EX上进行,测量的射线为6 MV光子线,该机器的剂量输出特性采用美国医学物理师协会第51号报告中的规范预先进行了校准[3]。将IMRT验证体模置于加速器治疗床上,十字线对准体模中心,利用激光灯和数字测距尺进行精确摆位。于测量点处放置FC65-G指形电离室,连接电缆至剂量仪DOSE-1,测量前给予预照射100 cGy,使其达到电离平衡,预照射后进行温度、气压校正和本底测量。每一计划均重复测量3次,测量结果取平均值。
同时,针对12例鼻咽癌IMRT计划,利用MatriXX进行平面相对剂量验证,将鼻咽癌IMRT计划的射野归一到0°机架角度,按设计进行照射、测量。利用二维电离室矩阵的附带软件对测量与计算的平面剂量进行比较,采用γ指数值的复合判断方法,以剂量点距离3 mm、相差3%为条件,以相对剂量方式进行,剂量以射野高剂量坪区归一,0≤γ指数≤1的点所占比例为通过率。为了避免低剂量区测量和计算不准确的影响,以及射野外测量点不参与比较的考虑,设置实测最高剂量点的10%做为阈值,测量点实测剂量低于此值者不进行比较。
-
在治疗计划系统中勾画出IMRT验证体模材料,由计划系统计算出其平均CT值。将验证体模的已知物理密度与平均CT值输入计划系统,对CT密度转换曲线进行修正,使得计划系统中验证体模图像的物理密度修正为对应的实际物理密度,重新计算剂量。在保持机器跳数、子野数、射野分布、子野分布等计划参数一致的前提下,比较前后两次计划计算值与实测值之间的差异,数据利用SPSS 16.0进行统计分析,修正前后之间的误差比较采用配对t检验方法,以P < 0.05为差异有统计学意义。
-
由电子密度体模各等效组织材料的平均CT值与对应的已知物理密度值输入计划系统Pinacle后绘制出的CT密度转换曲线见图 2。依据CT密度曲线,插值计算出的IMRT验证体模的物理密度约为1.020 g/cm3,而厂家说明书上提供的标准密度为1.045 g/cm3,二者约有0.025 g/cm3的差异。
参照多数医院使用二维电离室矩阵验证IMRT计划的方法,以γ指数通过率 > 90%为照射野剂量分布验证通过,则12例鼻咽癌IMRT计划验证测得的γ指数通过率为95.34%±1.50%,所有计划的通过率均 > 90%。由于IMRT平面相对剂量的验证不涉及剂量计算上具体的绝对值,因此,验证体模材料的微小密度差异并未给平面相对剂量的验证带来误差。
12例鼻咽癌IMRT计划分别按计划条件在加速器上模拟治疗方式的实际出束照射进行测量,将测量点处的计划剂量值与测得的靶区剂量值进行比较,修正前和修正后的百分相对误差全部控制在4%以内,符合临床要求。表 2给出了12例鼻咽癌IMRT计划前后两次计算值与实测值差异的统计分析,其中,误差=(计算值-测量值)÷测量值× 100%。由表 2数据可知,在同一个计划中,修正后的计算值比修正前低3 cGy左右,修正后的计算值与测量结果比较接近,误差控制在±2%以内,而修正前的计算值与测量结果的误差在0.79%~3.24%之间,修正后的误差显著低于修正前(t=19.256, P < 0.05)。12例鼻咽癌IMRT计划中,前后两次计算值与实测值的平均误差分别为1.96% ± 0.87%和0.63% ± 0.74%,最大误差分别为3.24%和1.81%。IMRT计划中验证体模微小的密度差异0.025 g/cm3,导致了绝对剂量验证中出现较大的偏差。
实测值(cGy) 计算值[cGy (误差,%)] 修正前 修正后 例1 210.2 217(3.24) 214(1.81) 例2 207.4 211(1.74) 208(0.29) 例3 203.4 205(0.79) 202(-0.69) 例4 194.2 196(0.93) 194(-0.10) 例5 202.4 206(1.78) 204(0.79) 例6 203.1 205(0.94) 203 (-0.05) 例7 200.2 205(2.40) 202(0.90) 例8 199.9 204(2.05) 201(0.55) 例9 196.8 203(3.15) 200(1.63) 例10 193.6 196(1.24) 194(0.21) 例11 192.1 197(2.55) 194(0.99) 例12 202.5 208(2.72) 205(1.23) 表 2 12例鼻咽癌IMRT计划前后两次计算值与实测值的比较分析
调强适形放射治疗剂量验证中CT密度值的校准分析
Calibration of CT density values in dosimetry verification of intensity modulated radiation therapy
-
摘要:
目的 基于调强适形放射治疗(IMRT)验证体模,分析放射治疗计划系统中CT密度转换曲线在IMRT剂量验证中的影响,并加以校准。 方法 利用CT电子密度体模刻度计划系统中的CT密度转换曲线,取12例鼻咽癌患者的IMRT计划,分别移植至IMRT验证体模,计算剂量分布,并于剂量均匀处放置测量点,利用电离室测量出其实际剂量。将IMRT验证体模的物理密度与对应的CT值输入计划系统,对CT密度转换曲线进行修正,重新计算剂量,在保持其他计划参数一致的前提下,比较前后两次计算值与实测值的差异。 结果 12例鼻咽癌IMRT计划中,前后两次计算值和实测值的平均误差分别为1.96%±0.87%和0.63%±0.74%,修正后的计算值误差控制在±2%以内,而修正前计算值的最大误差为3.24%。 结论 利用修正后的CT密度转换曲线,计划中的剂量计算值更加贴近实际测量值。为提高IMRT剂量验证的精确性,验证体模在使用之前必须进行CT密度值的校准、确认。 Abstract:Objective Based on intensity modulated radiation therapy(IMRT) phantom, the impact of CT-to-density conversion curve on dosimetry verification of IMRT is investigated and calibrated. Methods The electron density phantom was used to establish the CT-to-density conversion curve in radiation treatment planning system. IMRT plans of 12 nasopharynx carcinoma patients were chosen, copied to IMRT phantom and computed for the dose distribution. For each plan a measured point was put at the place where the dose was well-distributed and its dose value was measured using the ionization chamber. The physical density of IMRT phantom and its CT value were input into the planning system, to make a calibration for the CT-to-density conversion curve. The dose distribution was recomputed for each IMRT plan. Other parameters were kept the same in the plans and the differences between the computed dose values before and after correction were compared with the measured values. Results In 12 nasopharynx carcinoma IMRT plans, the average error of computed dose values was 1.96% ± 0.87% before correction and 0.63% ± 0.74% after correction, compared with measured values. The error between measured values and computed values after correction was less than ±2% whereas the maximum error of computed values before correction was 3.24%. Conclusions The com-puted dose values are closer to the measured values when using the calibrated CT-to-density conversion curve. The CT density values of IMRT phantom should be verified before usage, so as to increase the accuracy of IMRT dosimetry verification. -
表 1 美国CIRS公司Model 062电子密度体模的密度参数表
物理密度(g/cm3) 电子密度(x1023/m3) 相对水电子密度 体模(水等效) 1.01 3.346 1.002 肺(吸气) 0.20 0.634 0.190 肺(呼气) 0.50 1.632 0.489 脂肪 0.96 3.170 0.949 乳腺 0.99 3.261 0.976 肌肉 1.06 3.483 1.043 肝 1.07 3.516 1.052 骨(200mg/cm3) 1.16 3.730 1.117 致密骨(800 mg/cm3) 1.53 4.862 1.456 表 2 12例鼻咽癌IMRT计划前后两次计算值与实测值的比较分析
实测值(cGy) 计算值[cGy (误差,%)] 修正前 修正后 例1 210.2 217(3.24) 214(1.81) 例2 207.4 211(1.74) 208(0.29) 例3 203.4 205(0.79) 202(-0.69) 例4 194.2 196(0.93) 194(-0.10) 例5 202.4 206(1.78) 204(0.79) 例6 203.1 205(0.94) 203 (-0.05) 例7 200.2 205(2.40) 202(0.90) 例8 199.9 204(2.05) 201(0.55) 例9 196.8 203(3.15) 200(1.63) 例10 193.6 196(1.24) 194(0.21) 例11 192.1 197(2.55) 194(0.99) 例12 202.5 208(2.72) 205(1.23) -
[1] Cozzi L, Foqliata A, Buffa F, et al. Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapy. Radiother Oncol, 1998, 48(3): 335-338. doi: 10.1016/S0167-8140(98)00072-3 [2] 祁振宇, 黄劭敏, 邓小武. 放疗计划CT值的校准检测及其影响因素分析. 癌症, 2006, 25(1): 110-114.
[3] Tailor RC, Hanson WF, Ibbott GS. TG-51: experience from 150 institutions, common errors, and helpful hints. J Appl Clin Med Phys, 2003, 4 (2): 102-111. doi: 10.1120/jacmp.v4i2.2524 [4] 胡逸民. 肿瘤放射物理学. 北京: 原子能出版社, 1999.
[5] 杨克柽. 螺旋CT图像性能参数检测方法研究. 广州: 第一军医大学, 2002.
[6] Mutic S, Palta JR, Butker EK, et al. Quality assurance for computed-tomography simulators and the computed-tomography simulation process: report of the AAPM Radiation Therapy Committee Task Group No. 66. Med Phys, 2003, 30(10): 2762-2792. doi: 10.1118/1.1609271 [7] Beneventi S, Chionne F, Gobbi G, et al. Quantitative CT tomography for radiotherapy treatment planning: Calibration phantom and sources of error. Radiother Oncol, 1995, 37(suppl 1): S41. [8] Constantinou C, Harrington JC, Cadieux RA, et al. Dosimetry at lung-muscle and lung-bone interface using 6 MV and 10 MV X rays and 12-18 MeV electrons. Int J Radiat Oncol Biol Phys, 1988, 15(suppl 1): 244. [9] Constantinou C, Harrington JC, Dewerd LA. An electron density calibration phantom for CT-based treatment planning computers. Med Phys, 1992, 19(2): 325-327. doi: 10.1118/1.596862 [10] Mackie TR, El-Khatib E, Battista J, et al. Lung dose corrections for 6- and 15-MV x rays. Med Phys, 1985, 12(3): 327-332. doi: 10.1118/1.595691 [11] Waghorn B, Shah AP, Ngwa W, et al. A computational method for estimating the dosimetric effect of intra-fraction motion on step-and-shoot IMRT and compensator plans. Phys Med Biol, 2010, 55(14): 4187-4202. doi: 10.1088/0031-9155/55/14/015 [12] 邹华伟, 贾明轩, 吴荣, 等. 调强适形放射治疗计划剂量学的验证. 中华放射医学与防护杂志, 2004, 24(3): 256-258. doi: 10.3760/cma.j.issn.0254-5098.2004.03.025
[13] 张晓军, 王建华, 涂彧. 二维电离室矩阵在调强适形放疗剂量学验证中的应用. 国际放射医学核医学杂志, 2007, 31(5): 318-320. doi: 10.3760/cma.j.issn.1673-4114.2007.05.019