PET/CT稳定性检测数据分析及影响因素探讨

Data analysis and exploration of influencing factors in PET/CT stability testing

  • 摘要:
    目的 对Biograph Vision型PET/CT系统进行连续5年的长期稳定性能评估,系统分析性能漂移特征和关键影响因素,为临床影像质量控制体系建设提供循证依据。
    方法 参照《正电子发射断层成像(PET)设备质量控制检测标准》相关要求,严格遵循美国电气制造商协会NU 2—2012标准,于2021年10月至2026年3月对PET/CT系统探测器的工作状态、空间分辨率、灵敏度、等效噪声计数率(NECR)和散射分数、图像融合精度、标准摄取值(SUV)校准等6项核心性能指标开展连续性季度检测,累计完成14次标准化测试。性能指标的独立影响因素的筛选采用多重线性回归分析。
    结果 5年监测期间,PET/CT系统视野中心1 cm处横向、轴向半高宽的平均值分别为(3.62±0.04)、(3.78±0.05) mm,视野中心10 cm处横向、轴向半高宽的平均值分别为(4.26±0.06)、(4.46±0.06) mm。视野中心的5年平均灵敏度为(17.30±0.20) kcps/MBq。线源放射性活度浓度为30.3 kBq/ml时,NECR的峰值为245.0 kcps,对应的散射分数为33.8%。所有性能指标的年漂移率均<1.5%。图像融合精度平均配准误差为(1.60±0.10) mm。SUV校准结果显示,交叉校准校正系数的范围为0.970~1.030,平均值为1.000±0.020。以上检测项目14次检测结果的CV均≤2%,符合临床应用的阈值要求。多重线性回归分析显示,探测器的增益漂移(β=0.42,P<0.01)、机房温度波动(β=0.28,P<0.05)是影响PET/CT系统稳定性的独立危险因素。
    结论 Biograph Vision数字化PET/CT系统的长期稳定性良好,5年性能衰减程度低于行业平均水平。定期标准化的稳定性检测可有效预警早期的硬件故障,保障图像质量。

     

    Abstract:
    Objective  To evaluate the 5-year long-term stability of the Biograph Vision PET/CT system, systematically analyze its performance drift characteristics and key influencing factors, and provide evidence-based references for the construction of clinical image quality control systems.
    Methods According to Standard for Testing of Quality Control in Positron Emission Tomograph (PET) and National Electrical Manufacturers Association NU 2—2012 standard, quarterly serial tests covering six core performance indicators (detector operating status, spatial resolution, sensitivity, noise equivalent count rate (NECR) and scatter fraction, image fusion accuracy, and standardized uptake value (SUV) calibration) of PET/CT system were performed from October 2021 to March 2026, with a total of 14 rounds of standardized measurements. Multiple linear regression was used to identify independent influencing factors of performance indicators.
    Results During the 5-year monitoring period, the mean transverse and axial full width at half maximum at 1 cm from the field of view center were (3.62±0.04)  and (3.78±0.05) mm, respectively, whereas those at 10 cm from the field of view center were (4.26±0.06)  and (4.46±0.06)  mm, respectively. The average sensitivity at the field of view center was (17.30±0.20) kcps/MBq in 5 years. When the radioactivity concentration of the line source was 30.3 kBq/ml, the peak NECR was 245.0 kcps, with a scatter fraction of 33.8%. The annual drift rate of all performance indicators was less than 1.5%. The mean image fusion registration error was (1.60±0.10)  mm. The cross-calibration correction factor for SUV calibration ranged from 0.970 to 1.030, with a mean value of 1.000±0.020. The coefficient of variation of repeated measurements for all items was≤2%, meeting the clinical threshold requirements. Multiple linear regression analysis demonstrated that detector gain drift (β=0.42, P<0.01) and computer-room temperature fluctuation (β=0.28, P<0.05) were independent risk factors for PET/CT system stability.
    Conclusions The Biograph Vision digital PET/CT system exhibits favorable long-term stability, with lower 5-year performance degradation compared with the industry average. Regular standardized stability testing can effectively alert early hardware failures and guarantee image quality.

     

/