Abstract:
Objective To evaluate the 5-year long-term stability of the Biograph Vision PET/CT system, systematically analyze its performance drift characteristics and key influencing factors, and provide evidence-based references for the construction of clinical image quality control systems.
Methods According to Standard for Testing of Quality Control in Positron Emission Tomograph (PET) and National Electrical Manufacturers Association NU 2—2012 standard, quarterly serial tests covering six core performance indicators (detector operating status, spatial resolution, sensitivity, noise equivalent count rate (NECR) and scatter fraction, image fusion accuracy, and standardized uptake value (SUV) calibration) of PET/CT system were performed from October 2021 to March 2026, with a total of 14 rounds of standardized measurements. Multiple linear regression was used to identify independent influencing factors of performance indicators.
Results During the 5-year monitoring period, the mean transverse and axial full width at half maximum at 1 cm from the field of view center were (3.62±0.04) and (3.78±0.05) mm, respectively, whereas those at 10 cm from the field of view center were (4.26±0.06) and (4.46±0.06) mm, respectively. The average sensitivity at the field of view center was (17.30±0.20) kcps/MBq in 5 years. When the radioactivity concentration of the line source was 30.3 kBq/ml, the peak NECR was 245.0 kcps, with a scatter fraction of 33.8%. The annual drift rate of all performance indicators was less than 1.5%. The mean image fusion registration error was (1.60±0.10) mm. The cross-calibration correction factor for SUV calibration ranged from 0.970 to 1.030, with a mean value of 1.000±0.020. The coefficient of variation of repeated measurements for all items was≤2%, meeting the clinical threshold requirements. Multiple linear regression analysis demonstrated that detector gain drift (β=0.42, P<0.01) and computer-room temperature fluctuation (β=0.28, P<0.05) were independent risk factors for PET/CT system stability.
Conclusions The Biograph Vision digital PET/CT system exhibits favorable long-term stability, with lower 5-year performance degradation compared with the industry average. Regular standardized stability testing can effectively alert early hardware failures and guarantee image quality.