Abstract:
Objective To compare the effects of different monitoring and evaluation methods on individual dose monitoring results among interventional radiology workers and establish a scientific basis for improving individual dose monitoring strategies in this population.
Methods This was a retrospective cross-sectional study. The study analyzed individual dose monitoring data of 2109 person-times of interventional radiology workers monitored by Radioactive Testing Laboratory, Department of Occupational and Radiological Health, Yunnan Center for Disease Control and Prevention from 2021 to 2024, with each monitoring cycle lasting three months. Method 1 calculated the periodic individual dose equivalent Hp(10) ( Hp(10) refers to the monitoring of organs or tissues at a depth of 10 mm below the body surface and is used for effective dose assessment under specific conditions; hereinafter referred to as Hp(10)) using 0.1Ho (individual dose equivalent outside the lead apron). Method 2 calculated Hp(10) as 0.790Hu (individual dose equivalent under the lead apron)+0.051Ho. For Method 3, 0.790Hu+0.051Ho was adopted to calculate Hp(10) when Ho≥5Hu; under other circumstances, Hu was directly taken as Hp(10). The effective monitoring rate (the proportion of person-times with Hp(10) ≥ minimum detectable level) and Hp(10) levels of the three methods were compared. The non-parametric Kruskal-Wallis H test was used for multi-group comparison, and Dunn′s test (with Bonferroni correction) was adopted for pairwise comparisons.
Results The effective monitoring rates for Methods 1, 2, and 3 were 49.36%, 98.25%, and 97.72%, respectively, and their median Hp(10) values were 0.04, 0.11, and 0.13 mSv, respectively. The multi-group comparison showed statistically significant difference in Hp(10) calculated by different evaluation methods (H=833.897, P<0.01). Pairwise comparisons further revealed statistically significant differences in Hp(10) between all pairs of methods (Method 1 vs. Method 2: Z=−23.829; Method 1 vs. Method 3: Z=−28.029; Method 2 vs. Method 3: Z=−5.159; all P<0.01).
Conclusions Different monitoring and evaluation methods can substantially affect the individual dose monitoring results of interventional radiology workers. Therefore, the selection of appropriate monitoring and evaluation methods in routine practice is recommended to improve the reliability of monitoring results.